9.4 Triggering by Logic Signals (Logic Trigger)
161
9
Chapter 9 Trigger Settings
3. Trigger Pattern Make the settings of the logic trigger pattern.
Select
Setting Example Example 1: Trigger when the input signal matches any of the following cri-
teria:
Channel 1 (LA1): HIGH level
Channel 2 (LA2): LOW level
Trigger: OR
LA [1, 2, 3, 4]: [1 0 X X ]
Triggering occurs when the LA1 or LA2 trig-
ger criteria are met.
Example 2: Triggering occurs when the
input signal matches both of the following criteria:
Channel 1 (LA1): HIGH level
Channel 2 (LA2): LOW level
Trigger: AND
LA [1, 2, 3, 4]: [1 0 X X ]
X
Ignore signal. (default setting)
0
Trigger at LOW signal level.
1
Trigger at HIGH signal level.
To copy the setting to another channel
The Trigger settings window ([Logic Trig] sheet) can be used to copy a setting.
See:"8.9 Copying settings to other channels (calculation No.) (Copy function)" (
p.139)
Trigger Pattern
T
LA1 1
LA2 0
LA3 X
LA4 X
Trigger Pattern
T
LA1 1
LA2 0
LA3 X
LA4 X
T
If the conditions are met already when measurement is started (AND: all trig-
ger patterns are met, OR: one trigger pattern is met), triggering does not occur.
Triggering only occurs if the conditions are removed and then met again.
Triggers for standard logic channels (LA, LB, LC, and LD) are enabled regard-
less of the logic waveform display or unit type.