EasyManuals Logo

HP E4418B Programming Guide

HP E4418B
481 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
prog.book : chapter1.fm 82 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Status Reporting
1-82 HP E4418B/E4419B Programming Guide
Table 1-6: Bit Definitions - Standard Event Register
The standard event register is cleared when you:
send a *CLS (clear status) command.
query the event register using the *ESR? (event status register)
command.
The standard event enable register is cleared when you:
cycle the instrument power.
execute a *ESE 0 command.
Bit
Number
Decimal
Value
Definition
0 1 Operation Complete
All overlapped commands following an
*OPC
command have been completed.
1 2 Not Used. (Always set to 0.)
2 4 Query Error
A query error occurred, refer to error numbers 410 to
440 in the User’s Guide.
3 8 Device Error
A device error occurred, refer to error numbers 310 to
350 in the User’s Guide.
4 16 Execution Error
An execution error occurred, refer to error mumbers
211 to 241 in the User’s Guide.
5 32 Command Error
A command syntax error occurred, refer to error
numbers 101 to 161 in the User’s Guide.
6 64 Not Used. (Always set to 0.)
7 128 Power On
Power has been turned off and on since the last time
the event register was read or cleared.

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals