prog.book : ch7mem.fm 24 Mon Feb 15 09:24:37 1999
MEMory Subsystem
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
7-24 HP E4418B/E4419B Programming Guide
Error Messages
If more than 80 frequencies are in the list, error -108, “Parameter not
allowed” occurs.
If the frequencies are not entered in ascending order, error -220,
“Parameter error;Frequency list must be in ascending order” occurs.
If a table has not been specified using the MEMory:TABLe:SELect
command, the data cannot be entered into the table and error -221,
“Settings conflict” occurs.
If a frequency is sent which is outwith the allowed frequency range, error
-222, “Data out of range” occurs.