prog.book : ch9sens.fm 28 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1|2][:INPut][:MAGNitude]
9-28 HP E4418B/E4419B Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1|2][:INPut][:MAGNitude]? [MIN|MAX]
The query returns the current setting of the gain correction or the values
associated with MIN and MAX.
If [SENSe[1]]|SENSe2:CORRection:LOSS2 is used to set a gain level,
the result of this query is when the default unit is linear, and
when the default unit is logarithmic.
Query Example
Error Messages
If a channel offset value is entered using
[SENSe[1]]|SENSe2:CORRection:GAIN2 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, “Settings
Conflict” occurs. However, the channel offset value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not
automatically set ON.
The SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN1 command can
be used for the HP 8480 series power sensors when no sensor calibration
table has been set up. If a sensor calibration table is selected the error
-221, “Settings Conflict” occurs.
CORR:GAIN1? This command queries the current
setting of the calibration factor on
channel A.
1
Loss2
-----------------
Loss2

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals