Power Meter Remote Operation
Using Sensor Calibration Tables
HP E4418B/E4419B Programming Guide 1-35
Each string parameter returned indicates the name, type and size of a
stored sensor calibration table:
<string>,<type>,<size>
The <string>, <type> and <size> are all character data. The
<type> is always TABL. The <size> is displayed in bytes.
For example, a sample of the response may look like:
560,8020,“Sensor_1,TABL,220”,”Sensor_2,TABL,340” ....
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Hewlett-Packard Power Sensors (see Chapter 2,
“Editing Sensor Calibration Tables” in the User’s Guide). These power
sensors are:
DEFAULT
1
HP 8481A
HP 8482A
2
HP 8483A
HP 8481D
HP 8485A
R8486A
Q8486A
R8486D
HP 8487A
For further information on naming sensor calibration tables see “Naming
Sensor Calibration Tables”, on page 1-36.
1. DEFAULT is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
2. The HP 8482B and HP 8482H power sensors use the same data as the
HP 8482A.

Table of Contents

Other manuals for HP E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP E4418B and is the answer not in the manual?

HP E4418B Specifications

General IconGeneral
BrandHP
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals