Keysight N9915-90020 User’s Guide 235
SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
How to Set Up EMF Settings (Requires EMF (Option 358), GPS (Option 307), and SA
Mode (Option 233–Mixed Analyzers))
— Default Receiver side only antenna measurements
(Default – when no Tri Axial antenna is detected).
— Src/Rcvr Split to work with non-triaxial receiver and
source side antenna measurements.
— Triaxial XYZ (default, when Channel Power is on and
when a compatible triaxial antenna’s factors have
been imported) for measuring channel
measurements and EMF.
— Then <Back
to exit to the previous softkey menu.
e. Optional (Tri-Axial XYZ Antenna):
Verifying antenna and cable corrections are applied:
i. Press Scale/Amptd > More > Corrections
— Then verify the Apply Corrections Auto Disabled
softkey is set to Auto (This is the default setting when a
compatible USB antenna’s correction factors have been
imported, via the System softkeys).
— Then verify the X axis Antenna, Y axis Antenna, and
Z axis Antenna are all OFF. These can each be enabled
or disabled individually, but for EMF measurements all
three axes should be enabled (i.e., Antenna = Off, the
applications are setting the X/Y/Z).
— Then Press
Cable Off to open the cable correction
softkeys menu.
Verify Cable is ON. This can be disabled for applications
where an additional cable has not been connected to
the triaxial antenna. Before a cable can be used with
the EMF measurements it needs to have corrections
loaded. If you are going to use a cable, then choose as
needed:
—
Cable Off to enable a cable for use in your
antenna measurement.
— Edit Cable displays a cable corrections table
that you can edit
— New displays a cable corrections table that
you can edit with new cable correction
content.
Yes
– Choose Yes to overwrite the current cable
data
No
– Choose No to exit the cable data edit table
view