Keysight N9915-90020 User’s Guide 815
System SettingsFUTURE
Understanding the Fan Controls
Fan Error Messages
The FieldFox displays fan error messages, when the fan is not functioning
properly. Refer to Figure 31-11, Figure 31-12 on page 813, and Figure 31-13
on page 813.
Figure 31-13 Initial Fan Stuck Error Message – Displays For a Few Seconds
If the FieldFox displays a “FanStuck” or “Error: Internal Fan has stopped
spinning” message, to avoid incorrect measurements and possibly
damaging your FieldFox, shut your FieldFox down and contact Keysight.
Refer to
“Contacting Keysight” on page 5.
If the FieldFox displays a “Fan: Service” message the fan needs replacing.
Call Keysight and request a fan kit. Refer to
Figure 31-13 on page 813 and
to “Contacting Keysight” on page 5.