Keysight E4980A/AL Precision LCR Meter 65
Display Format
MEAS DISPLAY Page
When a DUT is connected to the measurement contacts, the setting current
value is different from the actual current passing through the DUT. For normal
measurement taken by applying the signal current through a device, refer to
“DC Bias Current Isolation” on page 97.
Figure 3-7 DC bias signal current
Setting limits for DC bias and test signal level
When both the DC bias and the test signal level are set under the following
conditions, the amount of DC bias plus the test signal level is limited to the
values listed in Table 3-19
• Option 001 is installed.
• The DC Bias key on the front panel is set to ON.
Table 3-18 DC bias and resolution (Option 001)
DC bias signal current level Resolution
(0A to 50 mA)
1 A
1
(50 mA to 100 mA) 10 A
1. Effective resolution is 3.3 A.
㪼㪋㪐㪏㪇㪸㫌㪼㪈㪇㪈㪎
㪈㪇㪇㱅
㪟㪠㪞㪟 㪣㪦㪮
㪛㪚㩷㪙㪠㪘㪪
㪚㪬㪩㪩㪜㪥㪫
㪪㪿㫆㫉㫋㪼㪻
Table 3-19 DC bias and test signal level-setting limits
Setting value Limit
DC bias Test signal
level
Vdc (V) Vosc (Vrms) Vosc 2 1.15 + Vdc 1.002 < 42 V
Vdc (V) Iosc (Arms) Iosc 2 115 + Vdc 1.002 < 42 V