Remote Commands
R&S
®
FSW
1018User Manual 1173.9411.02 ─ 43
<WindowType> Type of result display you want to use in the existing window.
See LAYout:ADD[:WINDow]? on page 1015 for a list of availa-
ble window types.
Example:
LAY:REPL:WIND '1',MTAB
Replaces the result display in window 1 with a marker table.
Usage: Setting only
LAYout:SPLitter <Index1>, <Index2>, <Position>
This command changes the position of a splitter and thus controls the size of the win-
dows on each side of the splitter.
Compared to the DISPlay[:WINDow<n>]:SIZE on page 1014 command, the
LAYout:SPLitter changes the size of all windows to either side of the splitter per-
manently, it does not just maximize a single window temporarily.
Note that windows must have a certain minimum size. If the position you define con-
flicts with the minimum size of any of the affected windows, the command will not work,
but does not return an error.
Figure 14-2: SmartGrid coordinates for remote control of the splitters
Setting parameters:
<Index1> The index of one window the splitter controls.
<Index2> The index of a window on the other side of the splitter.
Configuring the Result Display

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals