Remote Commands
R&S
®
FSW
1057User Manual 1173.9411.02 ─ 43
Example:
SENS:SWE:EGAT:AUTO MAN
SENS:SWE:EGAT:SOUR EXT2
SENS:SWE:EGAT:LEV:EXT2 1V
Sets the gating trigger to a level of 1 V at trigger port 2.
Manual operation: See "Gate Source Mode" on page 495
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt <Amount>
Defines the number of gate periods to be measured after a single trigger event.
Parameters:
<Amount> integer
Range: 1 to 1023
Increment: 1
*RST: 100
Example:
SWE:EGAT:CONT:PCO 50
Manual operation: See "Gate Period Count" on page 494
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth <Time>
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Parameters:
<Time> Range: 125 ns to 30 s
*RST: 5 ms
Default unit: S
Example:
SWE:EGAT:CONT:PLEN 10
Manual operation: See "Gate Period Length" on page 494
[SENSe:]SWEep:EGATe:CONTinuous[:STATe] <State>
Activates or deactivates continuous gating.
This setting is only available if [SENSe:]SWEep:EGATe is "On".
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0