Remote Commands
R&S
®
FSW
1068User Manual 1173.9411.02 ─ 43
This command is only available for instrument modelsR&S FSW43/50/67, for fre-
quency sweeps in the Spectrum application.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Preselector Adjust" on page 370
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S FSW after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the STAT:QUES:POW status register) and the
INPUT OVLD message in the status bar are cleared.
(See STATus:QUEStionable:POWer[:EVENt]? on page 1327 and "STA-
Tus:QUEStionable:POWer Register" on page 763).
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see "RF Input Protection" on page 367.
Suffix:
<ip>
.
1 | 2
For R&S FSW85 models with two RF input connectors:
1: Input 1 (1 mm [RF Input] connector)
2: Input 2 (1.85 mm [RF2 Input] connector)
For all other models:
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:CONNector <ConnType>
Determines which connector the input for the measurement is taken from.
Suffix:
<ip>
.
1 | 2
For R&S FSW85 models with two RF input connectors:
1: Input 1 (1 mm [RF Input] connector)
2: Input 2 (1.85 mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<ConnType> RF
RF input connector