Remote Commands
R&S
®
FSW
1091User Manual 1173.9411.02 ─ 43
Suffix:
<pb>
.
1..n
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF
Parameters:
<CMOffset> Offset of the mean voltage between the positive and negative
input terminal vs. ground
Range: -16 V to +16 V
Default unit: V
Manual operation: See "Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
set / " on page 372
[SENSe:]PROBe<pb>:SETup:DMOFfset
<DMOffset>
Sets the DM-mode offset. The setting is only available if a modular probe in DM-mode
is connected to the R&S FSW.
If the probe is disconnected, the DM-mode offset of the probe is reset to 0.0 V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S FSW I/Q Analyzer and I/Q Input User Manual.
Suffix:
<pb>
.
1..n
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF
Parameters:
<DMOffset> Voltage offset between the positive and negative input terminal
Default unit: V
Manual operation: See "Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
set / " on page 372
[SENSe:]PROBe<pb>:SETup:MODE <Mode>
Suffix:
<pb>
.
1..n
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF
Parameters:
<Mode> RSINgle | NOACtion
Setting Basic Measurement Parameters