Common Measurement Settings
R&S
®
FSW
372User Manual 1173.9411.02 ─ 43
Serial Number
Serial number of the probe
Remote command:
[SENSe:]PROBe<pb>:ID:SRNumber? on page 1090
Part Number
Rohde & Schwarz part number
Remote command:
[SENSe:]PROBe<pb>:ID:PARTnumber? on page 1089
Type
Type of probe:
Single-ended
Differential
Active Modular
Remote command:
[SENSe:]PROBe<pb>:SETup:TYPE? on page 1094
Mode
Mode for multi-mode modular probes. Determines which voltage is measured.
"DM-mode"
Voltage between the positive and negative input terminal
"CM-mode"
Mean voltage between the positive and negative input terminal vs.
ground
"P-mode"
Voltage between the positive input terminal and ground
"N-mode"
Voltage between the negative input terminal and ground
Remote command:
[SENSe:]PROBe<pb>:SETup:PMODe on page 1093
Common Mode Offset / Diff. Mode Offset / P Offset / N Offset /
Sets the offset for the probe, depending on the used mode (CM and DM mode both
use the "Common Mode Offset"). The setting is only available if a differential (R&S RT-
ZD) or modular (R&S RT-ZM) probe is connected to the R&S FSW.
If the probe is disconnected, the offset of the probe is reset to 0.0 V.
Note: If the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
Remote command:
[SENSe:]PROBe<pb>:SETup:CMOFfset on page 1090
[SENSe:]PROBe<pb>:SETup:DMOFfset on page 1091
[SENSe:]PROBe<pb>:SETup:NMOFfset on page 1092
[SENSe:]PROBe<pb>:SETup:PMOFfset on page 1093
Attenuation
Defines the attenuation applied to the input at the probe. This setting is only available
for modular probes.
"10:1"
Attenuation by 20 dB