Remote Commands
R&S
®
FSW
1103User Manual 1173.9411.02 ─ 43
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage: Setting only
Manual operation: See "Calibrate Reflection Short" on page 399
See "Calibrate Reflection Open" on page 399
[SENSe:]CORRection:METHod
<Type>
This command selects the type of measurement to be performed with the external gen-
erator.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe] on page 1098).
Parameters:
<Type> REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST: TRANsmission
Example:
CORR:METH TRAN
Sets the type of measurement to "transmission".
Manual operation: See "Calibrate Transmission" on page 399
See "Calibrate Reflection Short" on page 399
See "Calibrate Reflection Open" on page 399
[SENSe:]CORRection:RECall
This command restores the measurement configuration used for calibration.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe] on page 1098).
Example:
CORR:REC
Manual operation: See "Recall" on page 399
[SENSe:]CORRection[:STATe] <State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with [SENSe:]CORRection:COLLect[:ACQuire]
on page 1102.
Setting Basic Measurement Parameters