Common Measurement Settings
R&S
®
FSW
399User Manual 1173.9411.02 ─ 43
Save as Trd Factor......................................................................................................400
Reference Position......................................................................................................400
Reference Value..........................................................................................................400
Calibrate Transmission
Starts a transmission type measurement to determine a reference trace. This trace is
used to calculate the difference for the normalized values.
For details, see "Calibration Mechanism" on page 387.
Remote command:
[SENSe:]CORRection:METHod on page 1103
Calibrate Reflection Short
Starts a short-circuit reflection type measurement to determine a reference trace for
calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration trace is
calculated by averaging the two measurements. The order of the two calibration mea-
surements is irrelevant.
Remote command:
[SENSe:]CORRection:METHod on page 1103
Selects the reflection method.
[SENSe:]CORRection:COLLect[:ACQuire] on page 1102
Starts the sweep for short-circuit calibration.
Calibrate Reflection Open
Starts an open-circuit reflection type measurement to determine a reference trace for
calibration.
If both reflection-type calibrations (open circuit, short circuit) are carried out, the refer-
ence trace is calculated by averaging the two measurements. The order of the two cali-
bration measurements is irrelevant.
Remote command:
[SENSe:]CORRection:METHod on page 1103
Selects the reflection method.
[SENSe:]CORRection:COLLect[:ACQuire] on page 1102
Starts the sweep for open-circuit calibration.
Source Calibration Normalize
Switches the normalization of measurement results on or off. This function is only
available if the memory contains a reference trace, that is, after a calibration has been
performed.
For details on normalization, see "Normalization" on page 387.
Remote command:
[SENSe:]CORRection[:STATe] on page 1103
Recall
Restores the settings that were used during source calibration. This can be useful if
instrument settings were changed after calibration (e.g. center frequency, frequency
deviation, reference level, etc.).
Data Input and Output

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals