Common Measurement Settings
R&S
®
FSW
400User Manual 1173.9411.02 ─ 43
Remote command:
[SENSe:]CORRection:RECall on page 1103
Save as Trd Factor
Uses the normalized measurement data to generate a transducer factor. The trace
data is converted to a transducer with unit dB and stored in a file with the specified
name and the suffix .trd under
"C:\Program Files (x86)\Rohde-Schwarz\FSW\<version>\trd". The fre-
quency points are allocated in equidistant steps between start and stop frequency. The
generated transducer factor can be further adapted using the "Transducer" softkey in
the [SETUP] menu.
For more information on transducers, see Chapter 12.3.1, "Basics on Transducer Fac-
tors", on page 675.
This function is only available if Source Calibration Normalize is switched on.
Note: Note that the normalized measurement data is used, not the reference trace!
Thus, if you store the normalized trace directly after calibration, without changing any
settings, the transducer factor will be 0 dB for the entire span (by definition of the nor-
malized trace).
Remote command:
[SENSe:]CORRection:TRANsducer:GENerate on page 1104
Reference Position
Defines the position of the Result Frequency Stop in percent of the total y-axis range.
The top of the diagram is 100%, the bottom is 0%. By default, the 0 dB line is displayed
at the top of the diagram (100%).
This setting is only available if normalization is on (see "Source Calibration Normalize"
on page 399).
The reference line defined by the reference value and reference position is similar to
the Reference Level defined in the "Amplitude" settings. However, this reference line
only affects the y-axis scaling in the diagram, it has no effect on the expected input
power level or the hardware settings.
The normalized trace (0 dB directly after calibration) is displayed on this reference line,
indicated by a red line in the diagram. If you shift the reference line, the normalized
trace is shifted, as well.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition on page 1048
Reference Value
Defines the reference value to be displayed at the specified Result Frequency Start.
This setting can be used to shift the reference line and thus the normalized trace, simi-
lar to the Shifting the Display (Offset) defined in the "Amplitude" settings shifts the ref-
erence level in the display.
Shifting the normalized trace is useful, for example, to reflect an attenuation or gain
caused by the measured DUT. If you then zoom into the diagram around the normal-
ized trace, the measured trace still remains fully visible.
Data Input and Output

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals