Remote Commands
R&S
®
FSW
1160User Manual 1173.9411.02 ─ 43
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT <SearchLimit>
This command defines the left limit of the marker search range for all markers in all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<SearchLimit> The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST: left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation: See "Search Limits (Left / Right)" on page 216
See "Left Limit / Right Limit" on page 310
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt <SearchLimit>
This command defines the right limit of the marker search range for all markers in all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<Limit> The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST: right diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.