Remote Commands
R&S
®
FSW
1210User Manual 1173.9411.02 ─ 43
ON | 1
Switches the function on
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.
CALCulate<n>:FLINe<dl> <Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
Window
<dl> 1 to 4
frequency line
Parameters:
<Frequency> Note that you can not set a frequency line to a position that is
outside the current span.
Range: 0 Hz to Fmax
*RST: (STATe to OFF)
Default unit: HZ
Example:
CALC:FLIN2 120MHz
Sets frequency line 2 to a frequency of 120 MHz.
Manual operation: See "Vertical Line <x>" on page 558
CALCulate<n>:FLINe<dl>:STATe <State>
This command turns a frequency line on and off
Suffix:
<n>
.
Window
<dl> 1 to 4
frequency line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TLINe<dl> <Time>
This command defines the position of a time line.
Suffix:
<n>
.
Window
Analyzing Measurements (Basics)