Remote Commands
R&S
®
FSW
1214User Manual 1173.9411.02 ─ 43
CALCulate<n>:LIMit<li>:CONTrol:OFFSet <Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Offset> Numeric value.
The unit depends on the scale of the x-axis.
*RST: 0
Default unit: HZ
Manual operation: See "X-Offset" on page 565
CALCulate<n>:LIMit<li>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Distance> Numeric value.
The unit depends on the scale of the x-axis.
Default unit: HZ
Manual operation: See "Shift x" on page 569
CALCulate<n>:LIMit<li>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolMode> LINear | LOGarithmic
*RST: LIN
Example:
CALC:LIM:CONT:SPAC LIN
Manual operation: See "X-Axis" on page 568
Analyzing Measurements (Basics)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals