Remote Commands
R&S
®
FSW
1225User Manual 1173.9411.02 ─ 43
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Usage: Query only
Manual operation: See "Limit Check <n>" on page 254
See "Limit Check" on page 288
Programming Example: Using Limit Lines
The following examples demonstrate how to work with limit lines in a remote environ-
ment.
Example: Configuring Limit Lines........................................................................1225
Example: Performing a Limit Check....................................................................1226
Example: Configuring Limit Lines
This example demonstrates how to configure 2 limit lines - an upper and a lower limit -
for a measurement in a remote environment.
//------------- Configuing the limit lines ---------------------
CALC:LIM1:NAME 'FM1'
//Names limit line 1 'FM1'.
CALC:LIM1:CONT:MODE ABS
//Selects absolute scaling for the horizontal axis.
CALC:LIM1:CONT 1 MHz,50MHz,100 MHz,150MHz,200MHz
//Defines 5 horizontal definition points for limit line 1.
CALC:LIM1:UPP:MODE ABS
//Selects an absolute vertical scale for limit line 1.
CALC:LIM1:UNIT DBM
//Selects the unit dBm for limit line 1.
CALC:LIM1:UPP -10,-5,0,-5,-10
//Defines 5 definition points for limit line 1.
CALC:LIM1:UPP:MARG 5dB
//Defines an area of 5 dB around limit line 1 where limit check violations
//are still tolerated.
CALC:LIM1:UPP:SHIF -10DB
//Shifts the limit line 1 by -10 dB.
CALC:LIM1:UPP:OFFS -3dB
//Defines an additional -3 dB offset for limit line 1.
CALC:LIM3:NAME 'FM3'
//Names limit line 3 'FM3'.