Remote Commands
R&S
®
FSW
1236User Manual 1173.9411.02 ─ 43
MMEMory:SELect:CHANnel[:ITEM]:NONE
MMEMory:SELect[:ITEM]:NONE
This command does not include any of the following items when storing or loading a
configuration file.
●
Hardware configuration: MMEMory:SELect[:ITEM]:HWSettings on page 1235
●
Limit lines: MMEMory:SELect[:ITEM]:LINes:ALL on page 1235
●
Spectrogram data: MMEMory:SELect[:ITEM]:SGRam on page 1236
●
Trace data: MMEMory:SELect[:ITEM]:TRACe<1...3>[:ACTive]
on page 1236
●
Transducers: MMEMory:SELect[:ITEM]:TRANsducer:ALL on page 1237
Example:
MMEM:SEL:NONE
Manual operation: See "Items:" on page 631
MMEMory:SELect:CHANnel[:ITEM]:SPECtrogram <State>
MMEMory:SELect:CHANnel[:ITEM]:SGRam <State>
MMEMory:SELect[:ITEM]:SPECtrogram <State>
MMEMory:SELect[:ITEM]:SGRam <arg0>
This command includes or excludes spectrogram data when storing or loading a con-
figuration file.
Parameters:
<arg0> ON | OFF | 1 | 0
*RST: 0
Example:
MMEM:SEL:SGR ON
Adds the spectrogram data to the list of data subsets.
Manual operation: See "Items:" on page 631
MMEMory:SELect:CHANnel[:ITEM]:TRACe[:ACTive] <State>
MMEMory:SELect[:ITEM]:TRACe<1...3>[:ACTive] <arg0>
This command includes or excludes trace data when storing or loading a configuration
file.
Suffix:
<1...3>
.
irrelevant
Parameters:
<arg0> ON | OFF | 1 | 0
*RST: 0, i.e. no traces are stored
Example:
MMEM:SEL:TRAC ON
Manual operation: See "Items:" on page 631