Remote Commands
R&S
®
FSW
1239User Manual 1173.9411.02 ─ 43
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S FSW.
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument.
(Switch the [ON/OFF] key off and on).
MMEM:LOAD:STAT 1,'C:\R_S\INSTR\USER\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
See "Recall" on page 628
See "Recall in New Channel / Recall in Current Channel"
on page 632
MMEMory:LOAD:TYPE <arg0>
This command defines whether the channels that will be loaded with the subsequent
MMEM:LOAD:STAT command will replace the current channel or activate a new chan-
nel.
Parameters:
<arg0> NEW | REPLace
NEW
The loaded settings will be activated in a new channel.
REPLace
The loaded settings will replace the currently active channel.
*RST: NEW
Managing Settings and Results

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals