Remote Commands
R&S
®
FSW
1264User Manual 1173.9411.02 ─ 43
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage: Query only
Manual operation: See "Alignment Results:" on page 660
DIAGnostic:SERVice:INPut:MC:CFRequency <Frequency>
This command defines the frequency of the calibration signal for R&S FSW models 43
MHz and higher.
This command only takes effect if a microwave calibration signal is selected for input
(DIAGnostic:SERVice:INPut[:SELect] on page 1266)
Parameters:
<Frequency> *RST: 8.004 GHz
Default unit: Hz
Manual operation: See "Calibration Frequency MW" on page 716
DIAGnostic:SERVice:INPut:MC[:DISTance] <Bandwidth>
This command selects the distance of the peaks of the microwave calibration signal for
calibration of the YIG filter.
This command is only available for instrument models R&S FSW13/26.
Parameters:
<Bandwidth> WIDE | SMALl
SMALl
Small offset of combline frequencies.
WIDE
Wide offset of combline frequencies.
Manual operation: See "Calibration Frequency MW" on page 716
DIAGnostic:SERVice:INPut:PULSed:CFRequency <Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect] on page 1266.
Configuring the R&S
FSW