Remote Commands
R&S
®
FSW
1276User Manual 1173.9411.02 ─ 43
Suffix:
<si>
.
1..n
irrelevant
<fli> 1..n
Index in frequency response file list
Use [SENSe:]CORRection:FRESponse<si>:BASeband:
USER:FLISt<fli>:SIZE? to determine the maximum index
number.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Applies the data for magnitude results.
ON | 1
Does not apply the data for magnitude results.
*RST: 1
Example:
SENS:CORR:FRES:USER:FLIS2:MAGN:STAT ON
Manual operation: See "Magnitude" on page 695
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:PHASe[:
STATe] <State>
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:PHASe[:
STATe] <State>
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:PHASe[:STATe] <State>
Activates or deactivates the use of the correction data in the selected file for magnitude
results.
Suffix:
<si>
.
1..n
irrelevant
<fli> 1..n
Index in frequency response file list
Use [SENSe:]CORRection:FRESponse<si>:BASeband:
USER:FLISt<fli>:SIZE? to determine the maximum index
number.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Applies the data for magnitude results.
ON | 1
Does not apply the data for magnitude results.
*RST: 1
Example:
SENS:CORR:FRES:USER:FLIS2:PHAS:STAT ON
Manual operation: See "Phase" on page 695
Configuring the R&S FSW