Self-alignment results in secure user mode
Be sure to store self-alignment results before SecureUser Mode is enabled; see Chap-
ter 5.1.7, "Protecting Data Using the Secure User Mode", on page 43.
In secure user mode, the results are not stored permanently. Thus, if the currently
stored self-alignment results are not suitable, you must perform a self-alignment each
time you switch on the R&S FSW.
Start Self Alignment.................................................................................................... 659
Abort Self Alignment................................................................................................... 659
Await Warm-Up Operation before Self Alignment.......................................................660
Shut down Device after Self Alignment.......................................................................660
Reuse Old Alignment Data..........................................................................................660
Start Preselector Centering.........................................................................................660
Starting Touch Screen Alignment................................................................................660
Alignment Results:...................................................................................................... 660
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see Chapter 12.1.1, "Basics on Alignment", on page 656.
Note:
A running Sequencer operation is aborted when you start a self-alignment.
Remote command:
*CAL? on page 819, see also CALibration[:ALL] on page 1262
Abort Self Alignment
As long as the self-alignment data is being collected, the procedure can be canceled
using the "Abort Self Alignment" button.
Alignment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals