Remote Commands
R&S
®
FSW
826User Manual 1173.9411.02 ─ 43
Return values:
<ChannelType>,
<ChannelName>
For each channel, the command returns the channel type and
channel name (see tables below).
Tip: to change the channel name, use the INSTrument:
REName command.
Example:
INST:LIST?
Result for 3 channels:
'ADEM','Analog Demod','IQ','IQ
Analyzer','IQ','IQ Analyzer2'
Usage: Query only
Manual operation: See "Changing the Channel Name" on page 87
See "Selecting an application" on page 125
Table 14-2: Available channel types and default channel names in Signal and Spectrum Analyzer
mode
Application <ChannelType>
parameter
Default Channel name*)
Spectrum SANALYZER Spectrum
1xEV-DO BTS (R&S FSW-K84) BDO 1xEV-DO BTS
1xEV-DO MS (R&S FSW-K85) MDO 1xEV-DO MS
3GPP FDD BTS (R&S FSW-K72) BWCD 3G FDD BTS
3GPP FDD UE (R&S FSW-K73) MWCD 3G FDD UE
802.11ad (R&S FSW-K95) WIGIG 802.11ad
802.11ay (R&S FSW-K97) EDMG 802.11ay EDMG
Amplifier Measurements (R&S FSW-K18) AMPLifier Amplifier
Analog Demodulation (R&S FSW-K7) ADEM Analog Demod
Avionics (R&S FSW-K15) AVIonics Avionics
cdma2000 BTS (R&S FSW-K82) BC2K CDMA2000 BTS
cdma2000 MS (R&S FSW-K83) MC2K CDMA2000 MS
DOCSIS 3.1 (R&S FSW-K192/193) DOCSis DOCSIS 3.1
GSM (R&S FSW-K10) GSM GSM
I/Q Analyzer IQ IQ Analyzer
LTE (R&S FSW-K10x) LTE LTE
Multi-Carrier Group Delay (R&S FSW-K17) MCGD MC Group Delay
NB-IoT (R&S FSW-K106) NIOT NB-IoT
Noise (R&S FSW-K30) NOISE Noise
5G NR (R&S FSW-K144) NR5G 5G NR
*) If the specified name for a new channel already exists, the default name, extended by a sequential num-
ber, is used for the new channel.
Selecting the Operating Mode and Application

Table of Contents

Related product manuals