EasyManuals Logo

R&S FSW Series User Manual

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #828 background imageLoading...
Page #828 background image
Remote Commands
R&S
®
FSW
828User Manual 1173.9411.02 ─ 43
Setting parameters:
<ChannelName1> String containing the name of the channel you want to rename.
<ChannelName2> String containing the new channel name.
Note that you cannot assign an existing channel name to a new
channel; this will cause an error.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Example:
INST:REN 'IQAnalyzer2','IQAnalyzer3'
Renames the channel with the name 'IQAnalyzer2' to 'IQAna-
lyzer3'.
Usage: Setting only
Manual operation: See "Changing the Channel Name" on page 87
INSTrument[:SELect] <ChannelType> | <ChannelName>
This command activates a new channel with the defined channel type, or selects an
existing channel with the specified name.
Also see
INSTrument:CREate[:NEW] on page 824
Chapter 14.4.3, "Programming Example: Performing a Sequence of Measure-
ments", on page 831
Parameters:
<ChannelType> Channel type of the new channel.
For a list of available channel types see INSTrument:LIST?
on page 825.
<ChannelName> String containing the name of the channel.
Example:
INST IQ
Activates a channel for the I/Q Analyzer application (evaluation
mode).
INST 'MyIQSpectrum'
Selects the channel named 'MyIQSpectrum' (for example before
executing further commands for that channel).
Selecting the Operating Mode and Application

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals