Remote Commands
R&S
®
FSW
852User Manual 1173.9411.02 ─ 43
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?................................................856
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe...................................857
CALCulate<n>:LIMit<li>:ACPower[:STATe]....................................................................... 857
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute <LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the adjacent channels.
If you have defined an absolute limit as well as a relative limit, the R&S FSW uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li> irrelevant
Parameters:
<LowerLimit> The limit of the lower adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
Default unit: dBm
<UpperLimit> The limit of the upper adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
Default unit: dBm
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 188
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
This command turns the absolute limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
Absolute limit check for lower adjacent channel
*RST: 0
<State>
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 188