Remote Commands
R&S
®
FSW
857User Manual 1173.9411.02 ─ 43
Usage: Query only
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
<ch> 1..n
Alternate channel number
Parameters:
<State> ON | OFF | 1 | 0
Relative limit check for lower alternate channel
*RST: 0
<State>
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 188
CALCulate<n>:LIMit<li>:ACPower[:STATe] <State>
This command turns the limit check for ACLR measurements on and off.
In addition, limits must be defined and activated individually for each channel (see
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe,
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe,
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO]:ABSolute:STATe,
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO]:ACLR[:RELative]:
STATe and CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO][:
CACLr][:RELative]:STATe).
Suffix:
<n>
.
irrelevant
<li> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 182
See "Limit Checking" on page 188
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals