EasyManuals Logo

R&S FSW Series User Manual

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #854 background imageLoading...
Page #854 background image
Remote Commands
R&S
®
FSW
854User Manual 1173.9411.02 ─ 43
PASSED
Limit check has passed.
FAIL
Limit check has failed.
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage: Query only
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 188
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe
This command turns the relative limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
Relative limit check for lower adjacent channel
*RST: 0
<State>
Manual operation: See "Limit Check" on page 176
See "Limit Checking" on page 188
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute <LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S FSW uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<ch> 1..n
Alternate channel number
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals