Remote Commands
R&S
®
FSW
869User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
irrelevant
<li> irrelevant
<gap> 1 | 2
Gap (CACLR) channel number
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:LIM:ACP:GAP1:ABS:STAT ON
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO]:ACLR[:RELative]
<Limit>[,
<UpperLimit>]
This command defines the relative limit for the ACLR power in the specified gap chan-
nel. The reference value for the relative limit is the measured channel power.
If you define both an absolute limit and a relative limit, the R&S FSW uses the lower
value for the limit check.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<gap> 1 | 2
Gap (CACLR) channel number
Parameters:
<Limit> Defines the relative limit for the ACLR power in the specified gap
channel in dB.
Default unit: DB
<UpperLimit> Ignored.
Default unit: DB
Example:
CALC:LIM:ACP:GAP1:ACLR:REL 3dB
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:AUTO]:ACLR[:RELative]:STATe
<State>
This command turns the relative limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals