Remote Commands
R&S
®
FSW
874User Manual 1173.9411.02 ─ 43
<gap> 1 | 2
Gap (CACLR) channel number
Parameters:
<Limit> Defines the absolute limit of the specified gap channel.
Default unit: dBm
Parameters for setting and query:
<SBGaps> AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CALC:LIM:ACP:GAP2:MAN:LOW:ABS AB,44.2dBm
Example:
CALC:LIM:ACP:GAP:MAN:LOW:ABS? BC,DEF
Example: See "Example: Configuring and Performing an Asymmetrical
MSR ACLR Measurement" on page 895
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ABSolute:STATe
<SBGaps>, <State>
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>:MANual:LOWer:ABSolute:STATe?
<SBGaps>
This command turns the absolute limit check for the specified lower gap (CACLR)
channel on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
1..n
<li> 1..n
<gap> 1 | 2
Gap (CACLR) channel number
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Parameters for setting and query:
<SBGaps> AB | BC | CD | DE | EF | FG | GH
Name of the gap, defined by the letters of the surrounding sub
blocks (e.g. "AB" for the gap between sub blocks A and B).
Example:
CALC:LIM:ACP:GAP2:MAN:LOW:ABS:STAT BC,ON
Configuring and Performing Measurements