Remote Commands
R&S
®
FSW
909User Manual 1173.9411.02 ─ 43
Manual operation: See "Generator Frequency Coupling State" on page 232
CONFigure:GENerator:NPRatio:FREQuency:OFFSet <FrequencyOffset>
Defines a fixed offset to be applied to the generator frequency.
Parameters:
<FrequencyOffset> numeric value
Default unit: HZ
Example:
CONF:GEN:NPR:FREQ:OFFS 5
Manual operation: See "Frequency Offset" on page 233
CONFigure:GENerator:NPRatio:FREQuency[:FACTor]:DENominator
<Denominator>
Defines the denominator of the frequency-defining factor for the generator frequency
(see "f
Gen
" on page 232).
Parameters:
<Denominator> numeric value
Example:
CONF:GEN:NPR:FREQ:FACT:DEN 5
Manual operation: See "Denominator" on page 233
CONFigure:GENerator:NPRatio:FREQuency[:FACTor]:NUMerator <Numerator>
Defines the numerator of the frequency-defining factor for the generator frequency (see
"f
Gen
" on page 232).
Parameters:
<Numerator> numeric value
Example:
CONF:GEN:NPR:REQ:FACT:NUM 1
Manual operation: See "Numerator" on page 233
CONFigure:GENerator:NPRatio:NOTCh<notch>:BWIDth:ABSolute:CSTate?
Queries the state of the absolute notch bandwidth for the specified notch.
Suffix:
<notch>
.
1..n
Return values:
<ControlState> OFF | SUCCessful | ERRor
OFF
Signal generator control off
SUCCessful
Setting successfully applied on the signal generator
Configuring and Performing Measurements