EasyManuals Logo

R&S FSW Series User Manual

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #908 background imageLoading...
Page #908 background image
Remote Commands
R&S
®
FSW
908User Manual 1173.9411.02 ─ 43
OFF
Signal generator control off
SUCCessful
Setting successfully applied on the signal generator
ERRor
Control error, for example because a specified value cannot be
applied on the signal generator
Example:
CONF:GEN:NPR:EXT:ROSC:CST?
Usage: Query only
Manual operation: See "Reference Frequency" on page 230
CONFigure:GENerator:NPRatio:FREQuency:CENTer:CSTate?
Queries the state of the generator center frequency.
Return values:
<ControlState> OFF | SUCCessful | ERRor
OFF
Signal generator control off
SUCCessful
Setting successfully applied on the signal generator
ERRor
Control error, for example because a specified value cannot be
applied on the signal generator
Example:
CONF:GEN:NPR:FREQ:CENT:CST?
Usage: Query only
Manual operation: See "Frequency" on page 230
See "f
Gen
" on page 232
CONFigure:GENerator:NPRatio:FREQuency:COUPling[:STATe] <CouplingState>
Enables or disables frequency coupling between the R&S FSW and the connected
generator.
Parameters:
<CouplingState> ON | OFF | 0 | 1
OFF | 0
Switches the function off
A fixed frequency is used by the generator.
ON | 1
Switches the function on
The frequency defined on the analyzer is automatically also
used by the generator.
*RST: 0
Example:
CONF:GEN:NPR:FREQ:COUP:STAT ON
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals