Remote Commands
R&S
®
FSW
915User Manual 1173.9411.02 ─ 43
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
CONF:GEN:RFO:STAT ON
Manual operation: See "RF Output State" on page 229
CONFigure:GENerator:TARGet:PATH:BB?
Queries the BB signal path of the generator used for signal generation.
Return values:
<Path> A | B
Example:
CONF:GEN:TARG:PATH:BB?
Usage: Query only
Manual operation: See "Path RF/ Path BB" on page 229
CONFigure:GENerator:TARGet:PATH:RF <Path>
Selects the RF signal path of the generator to be used for signal generation.
Parameters:
<Path> A | B
Example:
CONF:GEN:TARG:PATH:RF B
Manual operation: See "Path RF/ Path BB" on page 229
CONFigure:SETTings:NPRatio
Usage:
Event
Manual operation: See "Query all Generator Setup Settings from Generator"
on page 231
CONFigure:SETTings:NPRatio:NOTCh
Usage:
Event
Manual operation: See "Query all Notch Settings from Generator" on page 226
14.5.6.4 Retrieving Results from an NP Measurement
CALCulate<n>:NPRatio:RESult?..................................................................................... 915
CALCulate<n>:NPRatio:RESult? <ResultType>
Queries the power results of the noise power ratio measurement.
Configuring and Performing Measurements