Remote Commands
R&S
®
FSW
931User Manual 1173.9411.02 ─ 43
For more information see "Relative limit line functions" on page 243.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STOP:FUNCtion on page 931).
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
For a detailed example see Chapter 14.5.7.11, "Example: SEM
Measurement", on page 950.
Manual operation: See "Rel Limit Start / Stop <n>" on page 254
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
<Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Function> OFF | MAX
Defines the function to be used to determine the relative limit
line stop value
MAX
The maximum of the relative and the absolute level is used as
the limit stop value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STOP and [SENSe:
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STOP:ABS commands to define these values.
Configuring and Performing Measurements