Remote Commands
R&S
®
FSW
949User Manual 1173.9411.02 ─ 43
Example:
CALC:ESP:PSE:AUTO OFF
Deactivates the list evaluation.
Manual operation: See "List Evaluation State (Result Summary)" on page 266
CALCulate<n>:ESPectrum:PSEarch[:IMMediate]
CALCulate<n>:ESPectrum:PEAKsearch[:IMMediate]
This command initiates a list evaluation.
Suffix:
<n>
.
Window
CALCulate<n>:ESPectrum:PSEarch:MARGin
<Threshold>
CALCulate<n>:ESPectrum:PEAKsearch:MARGin <Margin>
This command defines the threshold of the list evaluation.
Suffix:
<n>
.
Window
Parameters:
<Margin> Range: -200 to 200
*RST: 200
Default unit: dB
Example:
CALC:ESP:PSE:MARG 100
Sets the margin to 100 dB.
Manual operation: See "Margin" on page 266
CALCulate<n>:ESPectrum:PSEarch:PSHow <State>
CALCulate<n>:ESPectrum:PEAKsearch:PSHow <State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:ESP:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation: See "Show Peaks" on page 266
Configuring and Performing Measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals