Remote Commands
R&S
®
FSW
958User Manual 1173.9411.02 ─ 43
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
LIST:RANG2:INP:ATT:AUTO ON
Manual operation: See "RF Attenuation Mode" on page 287
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe <State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
The gain level is defined by [SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
on page 958.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
LIST:RANG2:INP:GAIN:STAT ON
Manual operation: See "Preamp" on page 287
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue] <Gain>
This command selects the preamplification level for the range.
The command requires option R&S FSW-B24.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Gain> 15 dB | 30 dB
The availability of preamplification levels depends on the
R&S FSW model.
R&S FSW8/13: 15dB and 30 dB
R&S FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST: OFF
Example:
LIST:RANG2:INP:GAIN 15
Configuring and Performing Measurements