Remote Commands
R&S
®
FSW
985User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
Window
<m> Marker
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:PHOLd:RESult?
This command queries the maximum standard deviation of the time domain power.
The query is only possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 835.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:RESult?
This command queries the standard deviation of the time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 835.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<StandardDeviation> Standard deviation of the signal during the measurement time.
Usage: Query only
14.5.10.4 Programming Example: Time Domain Power
This programming example demonstrates the measurement example described in
Chapter 7.9.6, "Measurement Example", on page 311 in a remote environment.
Configuring and Performing Measurements