Chapter 14—Maintaining the Library Calibrating the Touch Screen
August 2017 User GuideSpectra T50e Library
433
CALIBRATING THE TOUCH SCREEN
Ifthetouchpointsonthelibrary’stouchscreendonotproperlyalignwith
theuserinterfacegraphics,performthefollowingproceduretorecalibrate
thepositionofthetouchpoints.
Note: Youmustperformthisprocedurewhilestandinginfrontofthe
touchscreen.Theutilityisnotavailablewhenaccessingthe
librarythroughtheBlueScalewebinterface.
1. Logintothelibraryasauserwithsuperuseroradministrator
privileges.
2. ClickMENU,thenselectMaintenance > Utilities.TheBasicUtilities
screendisplays.
3. SelectCalibrate Touch Screen,thenclickSelect.TheDescriptionscreen
showsadescriptionoftheutility.
Note: TheCalibrate Touch Screenutilityisnotavailableifyouare
accessingthelibraryusingtheBlueScalewebinterface.
4. ClickRun.Thetouchscreendisplaysaseriesofcalibrationtarget
points.
5. Usingthestylus,toucheachtargetasitdisplaystorecalibratethetouch
pointsonthescreen.
Note: Toucheachpointatastraightonangleasclosetothecenterof
thetargetaspossible.
6. Whenthecalibrationroutineiscomplete,theUtilitiesscreen
redisplays.
Figure 301 Select the Calibrate Touch Screen utility.