9/29/09 Using ID Setting Files 7-15
P+A - The Logic Output turns on when a Peak
exceedance event occurs or when an event occurs
on one of the A:D channels.
R+P+A -
The Logic Output turns on when either
an RMS or Peak exceedance event occurs or when
an event occurs on one of the A:D channels.
I+A -
The Logic Output turns on when each inter-
val begins or when an event occurs on one of the
A:D channels.
R+I+A -
The Logic Output turns on when an
RMS exceedance event occurs, when each interval
begins, or when an event occurs on one of the A:D
channels.
P+I+A -
The Logic Output turns on when a Peak
exceedance event occurs, when each interval
begins, or when an event occurs on one of the A:D
channels.
R+P+I+A -
The Logic Output turns on when
either an RMS or Peak exceedance event occurs,
when each interval begins, or when an event
occurs on one of the A:D channels.
R/S -
The Logic Output turns on when the instru-
ment is in run mode and turns off when the instru-
ment is in stop mode.
Excd - The Logic Output turns on when a valid
exceedance is detected and after it’s minimum
duration has been met.
Logic Output Time Setting -
(SSA, LOG, ISM)
The Logic Output Time sets the number of seconds that the
logic output will remain turned “ON” following a triggering
event. Accordingly, this setting can take on any integral
value within the following range:
(0)sec to (255)sec