9/29/09 Using ID Setting Files 7-43
Use the u and d keys to move
from one setting to the next. Then use
the
r key to change the selected
setting. The
l key is used to
return to previous menus.
The following is a table of the Define Report settings. Each
setting is described later in greater detail.
Data Report - (LOG, SSA) The Data Report setting either enables or disables the
printing of the data report. The data report contains all the
basic measurement parameters for both the overall and
current measurements. This setting provides the following
two options:
No - The data report will not print.
Ye s - The data report will print.
Setup Report - (LOG, SSA) The Setup Report setting either enables or disables the
printing of the setup report. The setup report shows each of
the parameters and how they are set. This setting provides
the following two options:
No - The data report will not print.
Ye s - The data report will print.
Menu Items Available Settings Options
Data Report No
Yes
Setup Report No
Yes
SPL Histogram No
Yes
Pk-1 Histogram No
Yes
Pk-2 Histogram No
Yes
Intv Report No
Short
Long
Hist Report No
Yes