7-24 824 Reference Manual 9/29/09
measurement is reset. A special flag is set with all
exceedance records whose L
MAX
is greater that this level.
This setting can take on any integer value within the
following range:
0 to 255 dB
PK Excd Lvl 1 - (LOG, SSA) The Pk Excd Lvl 1 sets the trigger level for the primary peak
detector (Pk-I). When the primary peak detector level
exceeds this threshold, the corresponding Pk-I exceedance
count is incremented once. You can view this count on the
Stats-b display. This count is associated with the overall
measurement data and is reset when the overall
measurement is reset.
NOTE: The data for this comes from the
instantaneous peak level, selected by the
Weight setting.
This setting can include any integer value within the
following range:
0 to 255 dB
Pk Excd Lvl 2 - (LOG, SSA) The Pk Excd Lvl 2 sets the trigger level for the secondary
peak detector (Pk-II). When the secondary peak detector
level exceeds this threshold, the corresponding Pk-II
exceedance count is incremented once. You can view this
count on the Stats-b display. This count is associated with
the overall measurement data and is reset when the overall
measurement is reset.
This setting can include any integer value within the
following range:
0 to 255dB
Hysteresis - (LOG, SSA) To prevent multiple triggering/counting of the same event, a
hysteresis value may be entered. The event is triggered when
a level goes over a corresponding trigger level, but does not
become re-armed until the level drops below the value of the
trigger level minus the hysteresis value.