EasyManuals Logo

Larson Davis 824 User Manual

Larson Davis 824
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #430 background imageLoading...
Page #430 background image
12-22 824 Reference Manual 9/29/09
Command: "S 41,225.48"Response: ""// Key, unlock the
S227 command
Command: "S227,-7"Response: ""// Send LCAL, enter your
LCAL value
Command: "Q227"Response: " -7.00"// Verify LCAL,
verify your LCAL value
Command: "S 41,114"Response: ""// Restore Calibrator
Level
Here are typical Calibration Offset Levels (dB) for several
accelerometer sensitivities (re. 1mm/s2):
Sensitivity1mV/g10 mV/g100 mV/g1 V/g10 V/g
Cal. Offset23.00 dB13.00 dB-7.00 dB-27.00 dB-47.00 dB