12-22 824 Reference Manual 9/29/09
Command: "S 41,225.48"Response: ""// Key, unlock the
S227 command
Command: "S227,-7"Response: ""// Send LCAL, enter your
LCAL value
Command: "Q227"Response: " -7.00"// Verify LCAL,
verify your LCAL value
Command: "S 41,114"Response: ""// Restore Calibrator
Level
Here are typical Calibration Offset Levels (dB) for several
accelerometer sensitivities (re. 1mm/s2):
Sensitivity1mV/g10 mV/g100 mV/g1 V/g10 V/g
Cal. Offset23.00 dB13.00 dB-7.00 dB-27.00 dB-47.00 dB