Remote Commands
R&S
®
FSW
1046User Manual 1173.9411.02 ─ 43
*RST: 0
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 15
Switches on 15 dB preamplification.
Manual operation: See "Preamplifier" on page 455
INPut<ip>:GAIN[:VALue] <Gain>
This command selects the gain if the preamplifier is activated (INP:GAIN:STAT ON,
see INPut<ip>:GAIN:STATe on page 1045).
The command requires the additional preamplifier hardware option.
Suffix:
<ip>
.
1 | 2
For R&S FSW85 models with two RF input connectors:
1: Input 1 (1 mm [RF Input] connector)
2: Input 2 (1.85 mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<Gain> 15 dB | 30 dB
The availability of gain levels depends on the model of the
R&S FSW.
R&S FSW8/13/26: 15 dB and 30 dB
R&S FSW43 or higher: 30 dB
All other values are rounded to the nearest of these two.
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation: See "Preamplifier" on page 455
14.7.3.4 Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]................................................................. 1046
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE............................................. 1047
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE............................1047
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision.................................................. 1047
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition................................................. 1048
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing......................................1048
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe] <Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing.
Setting Basic Measurement Parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals