EasyManua.ls Logo

R&S FSW Series - Page 1078

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
1078User Manual 1173.9411.02 ─ 43
Example:
MIX ON
Activates the external mixer.
MIX:FREQ:HAND 78.0299GHz
Sets the handover frequency to 78.0299 GHz.
Manual operation: See "Handover Freq" on page 421
[SENSe:]MIXer<x>:FREQuency:STARt <Freq>
This command sets or queries the frequency at which the external mixer band starts.
Suffix:
<x>
.
irrelevant
Parameters:
<Freq> numeric value
Example:
MIX:FREQ:STAR?
Queries the start frequency of the band.
Manual operation: See "RF Start / RF Stop" on page 420
[SENSe:]MIXer<x>:FREQuency:STOP <Freq>
This command sets or queries the frequency at which the external mixer band stops.
Suffix:
<x>
.
irrelevant
Parameters:
<Freq> numeric value
Example:
MIX:FREQ:STOP?
Queries the stop frequency of the band.
Manual operation: See "RF Start / RF Stop" on page 420
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
This command restores the preset frequency ranges for the selected standard wave-
guide band.
Note: Changes to the band and mixer settings are maintained even after using the
[PRESET] function. Use this command to restore the predefined band ranges.
Suffix:
<x>
.
irrelevant
Example:
MIX:HARM:BAND:PRES
Presets the selected waveguide band.
Usage: Event
Manual operation: See "Preset Band" on page 421
Setting Basic Measurement Parameters

Table of Contents

Related product manuals