Remote Commands
R&S
®
FSW
1078User Manual 1173.9411.02 ─ 43
Example:
MIX ON
Activates the external mixer.
MIX:FREQ:HAND 78.0299GHz
Sets the handover frequency to 78.0299 GHz.
Manual operation: See "Handover Freq" on page 421
[SENSe:]MIXer<x>:FREQuency:STARt <Freq>
This command sets or queries the frequency at which the external mixer band starts.
Suffix:
<x>
.
irrelevant
Parameters:
<Freq> numeric value
Example:
MIX:FREQ:STAR?
Queries the start frequency of the band.
Manual operation: See "RF Start / RF Stop" on page 420
[SENSe:]MIXer<x>:FREQuency:STOP <Freq>
This command sets or queries the frequency at which the external mixer band stops.
Suffix:
<x>
.
irrelevant
Parameters:
<Freq> numeric value
Example:
MIX:FREQ:STOP?
Queries the stop frequency of the band.
Manual operation: See "RF Start / RF Stop" on page 420
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
This command restores the preset frequency ranges for the selected standard wave-
guide band.
Note: Changes to the band and mixer settings are maintained even after using the
[PRESET] function. Use this command to restore the predefined band ranges.
Suffix:
<x>
.
irrelevant
Example:
MIX:HARM:BAND:PRES
Presets the selected waveguide band.
Usage: Event
Manual operation: See "Preset Band" on page 421
Setting Basic Measurement Parameters