Remote Commands
R&S
®
FSW
1077User Manual 1173.9411.02 ─ 43
[SENSe:]MIXer<x>:THReshold <Value>
This command defines the maximum permissible level difference between test sweep
and reference sweep to be corrected during automatic comparison (see [SENSe:
]MIXer<x>:SIGNal on page 1076).
Suffix:
<x>
.
irrelevant
Parameters:
<Value> <numeric value>
Range: 0.1 dB to 100 dB
*RST: 10 dB
Example:
MIX:PORT 3
Manual operation: See "Auto ID Threshold" on page 424
Mixer Settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer<x>:FREQuency:HANDover....................................................................1077
[SENSe:]MIXer<x>:FREQuency:STARt.......................................................................... 1078
[SENSe:]MIXer<x>:FREQuency:STOP...........................................................................1078
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet................................................................ 1078
[SENSe:]MIXer<x>:HARMonic:BAND.............................................................................1079
[SENSe:]MIXer<x>:HARMonic:HIGH:STATe................................................................... 1079
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]................................................................. 1080
[SENSe:]MIXer<x>:HARMonic:TYPE............................................................................. 1080
[SENSe:]MIXer<x>:HARMonic[:LOW]............................................................................ 1080
[SENSe:]MIXer<x>:LOSS:HIGH.................................................................................... 1081
[SENSe:]MIXer<x>:LOSS:TABLe:HIGH..........................................................................1081
[SENSe:]MIXer<x>:LOSS:TABLe[:LOW].........................................................................1081
[SENSe:]MIXer<x>:LOSS[:LOW]................................................................................... 1081
[SENSe:]MIXer<x>:PORTs............................................................................................1082
[SENSe:]MIXer<x>:RFOVerrange[:STATe]......................................................................1082
[SENSe:]MIXer<x>:FREQuency:HANDover <Frequency>
This command defines the frequency at which the mixer switches from one range to
the next (if two different ranges are selected). The handover frequency for each band
can be selected freely within the overlapping frequency range.
This command is only available if the external mixer is active (see [SENSe:
]MIXer<x>[:STATe] on page 1075).
Suffix:
<x>
.
irrelevant
Parameters:
<Frequency> numeric value
Setting Basic Measurement Parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals