Remote Commands
R&S
®
FSW
1185User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
Window
<m> irrelevant
Example:
CALC:MARK:FUNC:NOIS:AOFF
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 835.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<NoiseLevel> Current noise level. The unit is the one currently active.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK2:FUNC:NOIS ON
Switches on noise measurement for marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:FUNC:NOIS:RES?
Outputs the noise result of marker 2.
Usage: Query only
Manual operation: See "Noise Measurement State" on page 536
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe] <State>
This command turns the noise measurement at the marker position on and off.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Analyzing Measurements (Basics)