Remote Commands
R&S
®
FSW
1200User Manual 1173.9411.02 ─ 43
For selective demodulation, the R&S FSW turns on a video trigger whose level corre-
ponds to the squelch level. Therefore it turns other triggers or gates off.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DEM:SQU ON
Signals below the level threshold are not sent to the audio out-
put.
Manual operation: See "Squelch" on page 547
14.8.3.15 Programming Examples for Using Markers and Marker Functions
Various programming examples on how to use markers and the special marker func-
tions are provided here.
The use of spectrogram markers is demonstrated in Chapter 14.8.2.7, "Programming
Example: Configuring a Spectrogram", on page 1148.
● Example: Basic Markers..................................................................................... 1200
● Example: Marker Search in Spectrograms......................................................... 1202
● Basic Frequency Sweep Measurement for Marker Function Examples............. 1203
● Example: Using a Fixed Reference Marker........................................................ 1203
● Example: Obtaining a Marker Peak List..............................................................1203
● Example: Measuring Noise Density.................................................................... 1204
● Example: Measuring Phase Noise......................................................................1204
● Example: Measuring the Power in a Channel Using Band Power Markers........ 1205
● Example: Measuring Characteristic Bandwidths (Using the n dB Down Marker)
............................................................................................................................ 1206
● Examples: Demodulating Marker Values and Providing Audio Output............... 1207
● Example: Performing a Highly Accurate Frequency Measurement Using the Signal
Count Marker...................................................................................................... 1208
Example: Basic Markers
This example demonstrates how to configure and define markers for a basic spectrum
measurement in a remote environment. It assumes that the basic frequency sweep
described in Chapter 14.15.1, "Programming Example: Performing a Basic Frequency
Sweep", on page 1375 has been performed and thus does not begin by presetting the
instrument.
//--------------Configuring marker behavior ------------------
DISP:MTAB ON
//Marker information is always displayed in a separate table.
Analyzing Measurements (Basics)