Remote Commands
R&S
®
FSW
1221User Manual 1173.9411.02 ─ 43
CALCulate<n>:LIMit<li>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Limit> Numeric value.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 1218.
*RST: -200
Default unit: dBm
Manual operation: See "Threshold" on page 567
Managing Limit Lines
Useful commands for managing limit lines described in the R&S FSW User Manual:
MMEM:SEL[:ITEM]:LIN:ALL
MMEM:STOR:TYPE
MMEM:LOAD:TYPE
Remote commands exclusive to managing limit lines:
CALCulate<n>:LIMit<li>:ACTive?...................................................................................1221
CALCulate<n>:LIMit<li>:COPY......................................................................................1222
CALCulate<n>:LIMit<li>:DELete.................................................................................... 1222
CALCulate<n>:LIMit<li>:STATe......................................................................................1222
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk....................................................................1223
MMEMory:LOAD<n>:LIMit............................................................................................ 1223
MMEMory:STORe<n>:LIMit.......................................................................................... 1223
CALCulate<n>:LIMit<li>:ACTive?
This command queries the names of all active limit lines.
Suffix:
<n>
.
irrelevant
<li> irrelevant
Return values:
<LimitLines> String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage: Query only
Analyzing Measurements (Basics)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals